Technical training of surgical products
2018-09-25 Views:7764

       In order to improve the management, service, knowledge, and skill level of staff and distributors, Kingsung will periodically organize training, including pre-job training, professional training etc., to better promote the product, we hold two sessions of advanced microsurgery knowledge training every year. All the lecturers are well-known experts and professors in the domestic microsurgery industry to share the benefits of Kingsung products to patients.


    2、Precautions for Coupler maintenance:

    A、The Anastomotic Instrument, Vessel Measuring Gauge,COUPLER Forceps, and Sterilization Tray must be sterilized prior to use;

    B、The Tray should not be sterilized at 132 degrees or higher. Microvascular Coupler and gauge can be sterilized at higher temperatures;

    C、The release button should be pressed when cleaning the handle, focusing on cleaning the gaps on both sides and the knob structure of the lower end, including the upper end and the small holes at the end of the handle;

    D、When cleaning micro forceps, the tip of forceps should be protected to avoid damage;

    E、After the Coupler is cleaned, rinse the equipment with the high-pressure water gun, especially the gap and knob, until there is no visible stain;

    F、During the operation, try to get into the habit of rinsing and soaking with heparin sodium salt or saline to keep the surface of stapler free of contaminants;

    G、The device should be cleaned immediately after use. If it cannot be cleaned in time, it should be put into a container with sterile distilled water to prevent the contaminants on the device from drying and adhering and reduce the service life;

    H、Microscopic surgical instruments are precision instruments and are recommended to be cleaned separately from orthopedic instruments to avoid damage to the device。